The Direct Cell Contact Test allows qualitative assessment of the cytotoxic potential of a test item according to ISO 10993-5. The test is performed by placing the test item on a monolayer of L929 cells. Cytotoxic substances released from the test item might cause deformation or even lysis of the cells beneath or around the test item, which can be evaluated microscopically.
Protocol |
|
Cell line |
L929 cells (ATCC No. CCL1, NCTC clone 929 (connective tissue mouse), clone of strain L (DSMZ)) |
Analysis |
Microscopic evaluation |
Incubation time |
24 h at 37 ± 1°C |
Quality controls |
Solvent Control: DMEM 10% FBS Negative control: High-density polyethylene material (HDPE) Positive control: 0.1% Zinc-diethyldithiocarbamate (ZDEC) |
Data delivery |
Cytotoxicity according to grading scores |
Positive prediction |
The test item meets the requirements of the test, if the response to the test item is not greater than grade 2 (mildly reactive) |